Venugopalan, L., Taylor, P.N., Cobb, J. E. and Swain, I. D., 2020. TetraGrip - a four channel upper limb FES device for people with C5/C6 tetraplegia: device design and clinical outcome. Journal of Medical Engineering & Technology, 44 (1), 38- 44.
Full text available as:
|
PDF
Lalitha & Cobb TetraGrip 2020.pdf - Accepted Version Available under License Creative Commons Attribution Non-commercial. 371kB | |
Copyright to original material in this document is with the original owner(s). Access to this content through BURO is granted on condition that you use it only for research, scholarly or other non-commercial purposes. If you wish to use it for any other purposes, you must contact BU via BURO@bournemouth.ac.uk. Any third party copyright material in this document remains the property of its respective owner(s). BU grants no licence for further use of that third party material. |
DOI: 10.1080/03091902.2020.1713239
Abstract
The TetraGrip is an inertial measurement unit-controlled surface upper limb FES device developed for improving hand functions of people with tetraplegia. The reliability of the control system and the repeatability and reproducibility of the device were assessed by analysing the results obtained when 14 able-bodied volunteers used the device. These volunteers were able to generate the control signals effectively once they had sufficient training. The two tetraplegic volunteers participated in a 12-week long clinical study (exercise, 4 weeks; functional tasks, 8 weeks), where they used the device to perform functional tasks. Outcome measures used were the grasp release test, the grip strength test, and the box and block test. Both tetraplegic volunteers showed improvement in performing the tasks specified in all outcome measures. The TetraGrip performed as intended when the able-bodied volunteers used it, and it improved the hand functions of both volunteers with tetraplegia. However, a larger clinical study is necessary to assess the performance of the device with a wider range of people with tetraplegia such as those with C5 complete/incomplete.
Item Type: | Article |
---|---|
ISSN: | 0309-1902 |
Uncontrolled Keywords: | Activities of daily living ; functional electrical stimulation ; inertial measurement unit ; outcome measures ; spinal cord injuries ; tetraplegia |
Group: | Faculty of Science & Technology |
ID Code: | 33346 |
Deposited By: | Symplectic RT2 |
Deposited On: | 03 Feb 2020 16:11 |
Last Modified: | 14 Mar 2022 14:19 |
Downloads
Downloads per month over past year
Repository Staff Only - |