Rolf, M.P., Hofman, M.B.M., Gatehouse, P.D., Markenroth-Bloch, K., Heymans, M.W., Ebbers, T., Graves, M.J., Totman, J. J., Werner, B., van Rossum, A.C., Kilner, P.J. and Heethaar, R.M., 2011. Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification--a multi-vendor study. Journal of Cardiovascular Magnetic Resonance, 13, 18 - ?.
Full text available as:
|
PDF
Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification--a multi-ve.pdf - Published Version Available under License Creative Commons Attribution Non-commercial. 669kB | |
Copyright to original material in this document is with the original owner(s). Access to this content through BURO is granted on condition that you use it only for research, scholarly or other non-commercial purposes. If you wish to use it for any other purposes, you must contact BU via BURO@bournemouth.ac.uk. Any third party copyright material in this document remains the property of its respective owner(s). BU grants no licence for further use of that third party material. |
Official URL: https://jcmr-online.biomedcentral.com/articles/10....
Abstract
PURPOSE: Eddy current induced velocity offsets are of concern for accuracy in cardiovascular magnetic resonance (CMR) volume flow quantification. However, currently known theoretical aspects of eddy current behavior have not led to effective guidelines for the optimization of flow quantification sequences. This study is aimed at identifying correlations between protocol parameters and the resulting velocity error in clinical CMR flow measurements in a multi-vendor study. METHODS: Nine 1.5T scanners of three different types/vendors were studied. Measurements were performed on a large stationary phantom. Starting from a clinical breath-hold flow protocol, several protocol parameters were varied. Acquisitions were made in three clinically relevant orientations. Additionally, a time delay between the bipolar gradient and read-out, asymmetric versus symmetric velocity encoding, and gradient amplitude and slew rate were studied in adapted sequences as exploratory measurements beyond the protocol. Image analysis determined the worst-case offset for a typical great-vessel flow measurement. RESULTS: The results showed a great variation in offset behavior among scanners (standard deviation among samples of 0.3, 0.4, and 0.9 cm/s for the three different scanner types), even for small changes in the protocol. Considering the absolute values, none of the tested protocol settings consistently reduced the velocity offsets below the critical level of 0.6 cm/s neither for all three orientations nor for all three scanner types. Using multilevel linear model analysis, oblique aortic and pulmonary slices showed systematic higher offsets than the transverse aortic slices (oblique aortic 0.6 cm/s, and pulmonary 1.8 cm/s higher than transverse aortic). The exploratory measurements beyond the protocol yielded some new leads for further sequence development towards reduction of velocity offsets; however those protocols were not always compatible with the time-constraints of breath-hold imaging and flow-related artefacts. CONCLUSIONS: This study showed that with current systems there was no generic protocol which resulted into acceptable flow offset values. Protocol optimization would have to be performed on a per scanner and per protocol basis. Proper optimization might make accurate (transverse) aortic flow quantification possible for most scanners. Pulmonary flow quantification would still need further (offline) correction.
Item Type: | Article |
---|---|
ISSN: | 1097-6647 |
Uncontrolled Keywords: | Aorta ; Blood Flow Velocity ; Equipment Design ; Europe ; Humans ; Image Interpretation, Computer-Assisted ; Magnetic Resonance Imaging ; Materials Testing ; Models, Cardiovascular ; Phantoms, Imaging ; Pulmonary Circulation ; Regional Blood Flow ; Reproducibility of Results |
Group: | Faculty of Health & Social Sciences |
ID Code: | 36364 |
Deposited By: | Symplectic RT2 |
Deposited On: | 10 Dec 2021 13:44 |
Last Modified: | 14 Mar 2022 14:31 |
Downloads
Downloads per month over past year
Repository Staff Only - |